Feature(may vary by option.)

●50Hz-100kHz (U2837)/ 50Hz-200kHz (U2836) , continuous frequency
●Basic accuracy: 0.05%, six digits resolution
●Support U-disk upgrade and rapid storage of test data to U-disk
●Real-time storage of all parameters without losing after shutdown
●Easy-used test interface
●Auto LCR function
●High speed and efficiency measurement: Max. 50meas / sec (with display)

[U2836]





Description

U2837U2837

General Introduction

  • U2837 & U2836 are whole new generation of general LCR meters with high-performance, beautiful appearance and easy operation. Providing stable 6 digit test resolution being equipped with a frequency of 50Hz-100kHz (U2837), 50Hz-200kHz (U2836), signal level of 10mV-1.0V (U2837), 10mV-2.0V (U2836) and max. test speed of 50meas/sec. These products can meet all requirements for measuring components and materials and provide guarant for production line quality control, incoming inspection and laboratory precision measurements.

Features:

  • 50Hz-100kHz (U2837),50Hz-200kHz (U2836), continuous frequency
  • Basic accuracy: 0.05%, six digits resolution
  • High speed and efficiency measurement: Max. 40meas / sec (with display)
  • Support U-disk upgrade and rapid storage of test data to U-disk
  • Real-time storage of all parameters without losing after shutdown
  • Easy-used test interface
  • Auto LCR function

General Technical Specifications

Test Parameter |Z|,C,L,R, X, |Y|,B,G,ESR,D,Q,θ
Basic Accuracy 0.05%
Measuring Speed Fast: 50, Med: 10, Slow: 2.5 (meas/sec)
Equivalent Circuit Series and Parallel
Ranging Mode Auto, Hold
Trigger Mode Internal, Manual, Auto, External and Bus
Correction Open / short
Display 480*272,4.3-inch TFT color screen
Memory Internal:100 groups,External:500 groups(U-disk)
Ls, Lp 0.00001μH~99.9999kH
Cs, Cp 0.00001pF~99.9999mF
R, Rs, Rp, X, Z 0.00001Ω~99.9999MΩ
G, Y, B 0.00001μS~99.9999S
ESR 0.00001mΩ~99.9999kΩ
D 0.00001~99.9999
Q 0.00001~99999.9
qr -3.14159~3.14159
qd -180.000°~180.000°
D% -99.9999%~999.999%
Comparators 5 bins, BIN1 ? BIN3,NG,AUX,PASS/FAIL LED display
Interface RS232C/USB-HOST/USB-CDC/ USB-TMC/HANDLER(optional)
Temperature & Humidity 0°C-40°C, ≤90%RH
Power Requirements 99V - 242V, 47.5Hz-63Hz
Power ≤ 20 VA
Dimension(W×H×D) 280mm×88mm×320mm
Weight approx. 2.5kg
U2837 small U2837 small
U2837 U2836
Test Frequency 50Hz ? 100kHz, 10mHz steps 50Hz ?200kHz, 10mHz steps
Output Impedance 30Ω, 50Ω, 100Ω 30Ω, 50Ω, 100Ω
Test Level 10mV - 1.0V, resolution:10mV 10mV - 2.0V, resolution:10mV

Whats in the box

  • LCR Meter, Four-terminal Kelvin test cable